GuidelinesBy device type2026.10.16
Blood Glucose Meter Imports into Korea: Separate Self-Testing, Professional Use and Strip Combinations
A strip-based blood glucose system cannot be reviewed through the meter alone. Connect users, specimens, meter and strip models, calibration, quality controls and storage with the proposed IVD scope.
Review the combined conditions of user, specimen, meter and strip. This guide addresses collected blood specimens and does not place continuous glucose monitoring systems in the same item category.
1. Separate self-testing from professional use
Home collection, measurement and interpretation by a lay user differ from professional operation in a clinical setting. Training assumptions and the ability to detect errors differ. Registered users and actual workflow matter more than online versus hospital distribution. Record user and location separately and compare them with the Korean plan.
If one model supports both environments, locate the evidence for each. Professional performance testing does not automatically establish that lay users can provide the correct sample volume or understand errors. See the IVD guide for the framework, then determine the application scope from product information.
2. Build a meter-to-strip matrix
Similar names do not establish one analytical system. Review strip codes, reaction principle, specimen, calibration and software together. This is a practical evidence-request example.
| Axis | Information | Cross-check |
|---|---|---|
| Meter | Model, software, displayed units | Tested versus supplied product |
| Strip | Model, manufacturer, reaction principle | Verified versus marketed combination |
| Specimen | Type, collection site, volume | Instructions versus performance conditions |
| Calibration | Coding or other applicable method | Meter, strip and operating instructions |
| Quality controls | Control material, conditions, interpretation | Quality checks and user information |
| Storage | Before/after opening, shelf life | Stability evidence and packaging |
List every proposed combination and separate unresolved ones from reviewed ones. A changed strip supplier requires reassessment even when the meter remains the same.
3. Read analytical performance with its conditions
For accuracy and precision, review specimen, range, comparison method, strip lots and meter identities. Align interference or specimen-characteristic limitations with Korean instructions. Confirm the required tests and acceptance criteria under the applicable product category and standards.
A report listing multiple lots and models does not, by itself, explain all sales combinations. Map results to the combinations they support. Distinguishing missing performance evidence from different storage conditions makes follow-up requests specific.
4. Check self-testing usability separately
MFDS's amendment published on 10 August 2026 establishes provisions for self-testing usability evidence, exemptions and evidence requirements. For a self-test, assess the amendment's supplementary provisions and applicability to the submission as well as analytical performance documentation.
Practical review topics include strip orientation, insufficient specimen, code mismatch where coding applies, units, error messages and repeat-test instructions. Omit steps absent from the product. This checklist does not replace regulatory usability evidence; the evaluated interface must align with final Korean instructions. See the usability guide.
5. Include opened storage and control materials
Vials and individually wrapped strips can have different in-use exposure conditions. An unopened expiry date alone does not explain repeated opening at home. Check specified conditions and stability evidence and reconcile dates and wording across Korean labels and instructions.
Establish the designated role of control material in checking the system rather than interpreting it as a patient diagnostic result. Its code, scope, storage and interpretation should match the actual meter and strips. Do not invent one universal control frequency for every glucose meter.
6. Confirm launch scope by combination
For each combination, connect users, specimens, codes, performance, stability, usability and labelling. The IVD import and distribution licence guide separates business and product procedures. Maintain change controls for strips, software and calibration methods affected by manufacturer updates.
Bring instructions, the model matrix and available performance evidence to request a glucose system import review. Clare Partners can identify the product scope and evidence gaps first.
Official references: MFDS personal blood glucose system safe-use information · MFDS IVD review regulation amendment No. 2026-58. Checked 2026-09-28. Applicability dates and exemptions require review of the supplementary provisions and conditions. Tables and workflow are practical recommendations.
Frequently asked questions
- Q. Can a professional meter simply be sold for home use?
- A change in users and environment is more than a sales-channel change. Check registered intended use, specimen, performance and applicable self-testing usability evidence.
- Q. Can another manufacturer's strips be sold with the same meter?
- Do not assume compatibility. Verify the combination's registration scope, performance evidence, calibration method, instructions and change responsibilities.
- Q. Does this guide also cover continuous glucose monitors?
- This guide concerns collected specimens analyzed with strips and a meter. Continuous systems using in-body sensors have different designs and require separate classification and evidence assessment.