GuidelinesBy procedure2026.11.08
Veterinary Medical Device Test Report Requirements — Nine Issuing Bodies, Model Name and Raw Material Identity and the Three-Year Rule in One Table
A CB or KOLAS logo and a date of issue are not enough to judge whether a test report can be used for approval. Article 6(2) of the Regulation on the Review of Technical Documents, etc. for Veterinary Medical Devices sets three requirements for test data: one of nine issuing bodies; identity of model name and list of raw materials with the product applied for, and conformity with Annex 2; and additional data where three years have passed since issue as at the submission date. This guide sets out a table for checking the three requirements for each document, the route under Article 23 of the Approval Notice by which a designated test laboratory recognizes external reports, and the supplementation and rejection provisions.
Key takeaway — Test data submitted for a veterinary medical device approval review must meet three requirements at once. First, they must come from one of the nine issuing bodies listed in Article 6(2) of the Review Notice. Second, they must have the same model name and list of raw materials as the product for which review is applied, and contain the Annex 2 entry items. Third, if more than three years have passed since the date of issue as at the submission date, data confirming that the product has not changed since issue must be attached. The logo and date on the cover are only the starting point for these three, not the conclusion.
This article is based on the APQA notices Regulation on the Review of Technical Documents, etc. for Veterinary Medical Devices (No. 2020-18) and Regulation on the Approval, etc. of Veterinary Medical Devices (No. 2020-19), and the Rules on the Handling of Veterinary Pharmaceuticals, etc. (MAFRA Ordinance No. 741, in force 31 October 2025), as checked on 6 October 2026. For the overall framework of jurisdiction and classes, see the veterinary medical device registration guide, and for the full set of documents to request from the manufacturing source, see the Class 2 technical document request table.
Which applications these requirements apply to
Article 3(1) of the Review Notice defines the scope of the review of technical document data as veterinary medical devices for which manufacturing or import product approval (including change approval) is sought under Article 7(4) of the Handling Rules. In other words, products on the approval route are covered: Classes 2, 3 and 4, and Class 1 items that are not substantially equivalent to an item already approved or notified.
The attachments for the import notification of a substantially equivalent Class 1 item are set separately in Article 16(2)(ii) of the Handling Rules: documents on product information (Article 5(3)(ii)(a)) and certificates of manufacture and sale or documents that can prove the same; attachment of test data is not written into that provision. Whether the route is notification or approval must be settled first before you can tell whether the requirements in this article apply.
How two provisions interlock: the designated laboratory's report and recognition of external reports
On the approval application side there is one more provision. Article 5(3)(ii)(c) of the Handling Rules lists as an attachment to the product approval application a test report issued by a test laboratory designated by the Minister of Agriculture, Food and Rural Affairs under Article 51-3, and Article 16(1)(ii)(a) applies the same document to import approval. How this attachment connects to the supporting data requirements in Article 6(2) of the Review Notice is explained in Chapter 4 of the Approval Notice.
| Approval Notice provision | Content |
|---|---|
| Article 21 (application for testing) | When applying to a designated test laboratory for testing, attach the technical document data, for imports a copy of the customs documents evidencing the test device, and any test reports recognized under Article 23. |
| Article 22 (testing) | The head of the test laboratory conducts testing according to the test specifications and test items set in the technical documents, and may commission items it cannot perform to domestic or overseas public test institutions, accredited research institutions, universities and the like. Duplicate safety testing of raw materials or identical component parts within the same product group is omitted. |
| Article 23 (recognition of test reports) | Where the applicant submits a report issued by one of the following bodies, the head of the test laboratory may review and recognize it: 1. a veterinary medical device test laboratory designated by the Minister of Agriculture, Food and Rural Affairs; 2. a medical device test laboratory designated by the Commissioner of the Korea Food and Drug Administration (as worded in the provision); 3. in other cases, a body issuing test data under the Review Notice; 4. a body with a mutual recognition agreement with a body under point 1 or 2. Copies must have their reliability secured by confirmation of the issuing body or notarisation, etc.; recognition is limited to results obtained by test methods at or above those recognized by the APQA Commissioner; and where only some items meet the requirements, only those items are recognized and the rest are tested under Article 22. |
| Article 24 (issue of test reports) | The head of the test laboratory issues a test report in Form 5 as the result of testing conducted under Articles 22 and 23. |
Reading this structure reveals the practical sequence. CB data or overseas reports sent by the manufacturing source can be submitted to the designated test laboratory for recognition under Article 23; recognized items are carried over as they are, unrecognized items are tested by that laboratory, and one test report is issued. Which report will be recognized under which point must be confirmed with the designated test laboratory for each issuing body and test item; this article does not assert the recognition outcome of any specific body. The list and scope of designation of designated test laboratories were not checked for this article.
Requirement 1: nine issuing bodies
Article 6(2) provides that, among the supporting data under paragraph (1), test data must fall under one of the following points. The left column is the provision; the right column is what the person in charge checks in the actual document.
| Point | Provision (summary) | What to check in the document |
|---|---|---|
| 1 | Test data issued by a veterinary medical device test laboratory designated by the Minister of Agriculture, Food and Rural Affairs | Whether the issuing body is designated and its scope of designation |
| 2 | Test data issued by a medical device test laboratory designated by the Commissioner of the Korea Food and Drug Administration (as worded in the provision) | Whether the issuing body is MFDS-designated and whether the test item falls within the scope of designation |
| 3 | Test data issued by a national certification body (NCB) under the IECEE CB Scheme operated by the International Electrotechnical Commission (IEC) | Whether issued by an NCB; the distinction between the CB certificate and the test report; models and options included in the report |
| 4 | GLP test data issued by a GLP test facility accredited under the OECD Principles of Good Laboratory Practice | Scope of GLP accreditation and whether the test was performed under GLP |
| 5 | Data published in a specialist journal indexed in the Science Citation Index (SCI) | Whether the journal is SCI-indexed and whether the data connect to the product applied for |
| 6 | Data tested by a domestic or overseas specialist institution such as a university or research institute, issued by the head of that institution, which can be recognized as valid after review of the entries on the outline of test facilities, main equipment, composition of research staff and testers' research experience, etc.; where no domestically or internationally recognized test standard exists, safety and performance test data on design and development conducted under the manufacturer's quality management system | Whether issued by the head of the institution; whether facilities, equipment, staff and experience are stated; whether a recognized standard exists |
| 7 | All data submitted and evaluated at the time of approval in the country of development (clinical trial results, etc.), confirmed as received or approved by the government of the country of development or a registration body entrusted with approval work, or notarised copies thereof | Link between the approval record in the country of development and the data; confirmation or notarisation document |
| 8 | Test data issued by a veterinary medical device clinical trial institution designated by the APQA Commissioner | Designation status and the record of approval of the clinical trial plan |
| 9 | Test data issued by a veterinary medical device non-clinical test institution designated by the APQA Commissioner | Designation status and scope of testing |
KOLAS does not appear as a separate item in this list. So the description ‘KOLAS report’ alone does not connect directly to any point. The practical question is to check, body by body and test item by test item, whether the same body is also an MFDS-designated test laboratory under point 2, whether it can be reviewed under the specialist institution route in point 6, or whether it is a mutual recognition agreement body under Article 23(1)(iv) of the Approval Notice. Conversely, CB data under point 3 have only met the issuing body requirement and must pass the two requirements below separately.
Requirement 2: identical model name and list of raw materials, and conformity with Annex 2
The main text of Article 6(2) requires test data to have the same model name and list of raw materials as the product for which technical document review is applied, and to conform to Annex 2. Two comparisons are needed.
The first is the identity comparison. Place the test sample name and model name on the report side by side with the model name on the application, and compare the raw materials stated on the report (especially parts that contact or are inserted into the animal) with the list of raw materials on the application. If the overseas sales model name differs from the Korean application model name, manufacturer documents explaining the relationship are needed. The manufacturer answering ‘same series’ is different from the model name and raw materials being identical on paper.
The second is the Annex 2 comparison. Annex 2, ‘content to be included in data on measured values’, has twelve items.
| No. | Annex 2 item | Where to look in the report |
|---|---|---|
| 1 | Name and address of the test laboratory | Cover and header |
| 2 | Name, representative and address of the company requesting the test | Applicant field |
| 3 | Serial number of the test report and page numbers for each page and the total | Every page |
| 4 | Name and marking of the test sample | Sample identification field |
| 5 | Date of receipt or date of testing | Receipt and test records |
| 6 | Date of issue of the test report | Cover. The reference date for requirement 3 |
| 7 | Signature or seal of the person responsible for the test report | Signature block |
| 8 | Test criteria and test methods; where no standard exists, the reason for setting them | Criteria for each test item |
| 9 | Test results | Results table |
| 10 | Matters on sampling and method of the test sample (where separate pre-treatment is needed) | Sample preparation record |
| 11 | Test system information; number, sex, strain, species, age, etc. of test animals (for tests using animals, cells or micro-organisms) | Biological test report |
| 12 | Environmental factors of the test where they affect the results | Test conditions field |
A single certificate page contains hardly any of these items. The complete test report and its annexes are needed for the comparison, and an excerpt lacking page numbers, as in item 3, fails the requirement by itself.
Requirement 3: three years from the date of issue as at the submission date
The proviso to Article 6(2) requires that, for test data issued more than three years before the submission date, data confirming that the product has not changed since the date of issue be submitted in addition. The two dates for the calculation are the date of issue of the report (Annex 2 item 6) and the submission date of the review application. The date the data were received or the contract date is not the reference.
If the planned submission date slips during preparation, whether the three years have passed may change, so record the date of issue and the planned submission date together in the document list and recalculate just before filing.
Distinguish what the provision does and does not provide.
- What it provides: for data older than three years, ‘data confirming no change’ are submitted in addition.
- What it does not provide: a duty to retest, the format of the confirming data, or automatic acceptance of data less than three years old.
If there has actually been a change in the meantime, data confirming no change cannot be produced. In that case, review whether the changed part affects the existing test items; and where the change concerns the main technical characteristics in Annex 1 and has a significant effect on safety and effectiveness, the device becomes subject to review including safety and effectiveness data under Article 3(2)(v) of the Review Notice. Rather than producing a confirmation that papers over the change, the right order is to set out the scope of the change and the usability of the existing data.
A review table to fill in for each document
This is our own management table for checking the three requirements document by document. It is not a statutory form and is for preparation before filing.
| Document | Issuing body (which point of Article 6(2)) | Test items and applied standard | Model name and list of raw materials match | Three years from issue to planned submission | Missing Annex 2 items | Open question |
|---|---|---|---|---|---|---|
| (fictional example) Electrical safety test report | Point 3, NCB confirmation requested | IEC 60601-1 series, as stated in the report | Option B not in the report's model list | Issued 2023-02; past three years if submitted 2026-12 | No page numbers | Whether option B is covered; reissue of the complete report |
| (fictional example) Biological safety test | Point 4, GLP status being confirmed | Based on contacting part materials | Material name only for contacting parts; specification not stated | Not past | Confirm item 11 test system information | Compare the list of raw materials with the material names |
Three practices help when running the table. Do not leave the point number blank and move on with ‘it will probably be recognized’. A model name mismatch and the three-year rule are separate problems, so do not mix them in one cell. Split the open questions into questions for the manufacturing source, questions to confirm with the designated test laboratory, and questions to confirm with APQA.
What happens if a requirement is missing: supplementation and rejection
Article 7(1) of the Review Notice lists as a ground for a request for supplementation the case where the type, scope or requirements, etc. of the attached data do not conform to Articles 5 and 6. Supplementation may be requested only once, with a period of up to 60 days, and if it is not submitted within that period a reminder may be issued to supplement within 10 days (Article 7(2)). If it is still not submitted within the reminder period, or does not meet the requirements even after supplementation, that is a ground for rejection (Article 7(3)).
The product approval processing periods given by Government24 are 40 days in total for Class 1 and 2 requiring technical document review, 60 days in total for Class 3 and 4, and 80 days in total where a safety and effectiveness review is needed, with processing periods of 6 days or more calculated excluding Saturdays and public holidays (checked 6 October 2026). If a test data requirement is missing and supplementation is requested, the supplementation period is added to these. That is why the data are screened against the three requirements before filing.
Common misunderstandings
| Common explanation | What the provisions provide | What to check |
|---|---|---|
| A CB report is enough | Point 3 is the issuing body requirement. Identity, Annex 2 and the three-year rule are separate. | Model list, complete report, date of issue |
| KOLAS is never recognized | It is merely not a separate item; no wording denies recognition either. | Whether the issuing body falls under another point or Article 23(1)(iv) |
| After three years it is retesting | The provision adds data confirming no change. | Change history since the date of issue |
| Under three years, we use it as it is | Only the three-year requirement has been passed. | Model name and raw material identity, Annex 2 |
| Just get a new date of issue | A reissued date does not change the test sample. | Whether the test sample matches the current specification |
| We have an overseas certificate, so we have test data | Certificates and test data part ways at the Annex 2 items. | A report with test criteria, results and signature |
The general issues in using overseas reports resemble those in the human-use guide to using foreign test reports, but veterinary devices have different provisions and a different list of issuing bodies, so recheck against the tables in this article. For the order of requesting data from the manufacturing source, use the request table in the overseas manufacturer document request guide alongside.
Documents that speed up the review if prepared before inquiring
- Report cover and table of contents, and the pages listing models and options
- Issuing body information (name, country, documents presented as the basis of designation or accreditation)
- Planned application model name and list of raw materials (including contacting part materials)
- Date of issue of each report and the planned filing date
- Change history or confirmation of no change sent by the manufacturing source
Please mention veterinary medical devices in your message when you use the veterinary test data review inquiry. We will set out which point your documents fall under, what is missing on identity and the three-year rule, and what to ask the manufacturing source and the test laboratory respectively. Additional testing costs and agency fees are distinguished, and agency fees are quoted separately according to the scope of review and application. The scope of veterinary services is on the veterinary medical device services page.
Prepared and sources checked: 2026-10-06. Sources: Regulation on the Review of Technical Documents, etc. for Veterinary Medical Devices (APQA Notice No. 2020-18) Article 3, Article 6(1) and (2), Article 7, Annex 2 · Regulation on the Approval, etc. of Veterinary Medical Devices (Notice No. 2020-19) Articles 21 to 24 · Rules on the Handling of Veterinary Pharmaceuticals, etc. (MAFRA Ordinance No. 741, in force 31 October 2025) Article 5(3), Article 7(4), Article 16(1) and (2), Article 51-3 · Government24 veterinary medical device manufacturing (import) product approval processing period guidance. The provisions were checked against the original text on the Korean Law Information Center; the review table and fictional example are our own practical tools, not statutory forms. The list of designated test laboratories was not checked for this article.
Frequently asked questions
- Q. Can a report from a KOLAS-accredited laboratory not be used for veterinary medical device approval?
- KOLAS is not listed as a separate item among the nine issuing bodies in Article 6(2) of the Review Notice. That does not mean the provision says such a report cannot be used. Whether the issuing body falls under another point of the same paragraph, for example an MFDS-designated medical device test laboratory or the validity review route for specialist institution data, or under the mutual recognition agreement bodies in Article 23(1)(iv) of the Approval Notice, must be checked body by body and test item by test item.
- Q. If we have a CB report, is the electrical safety data done?
- Article 6(2)(iii) covers test data issued by a national certification body (NCB) under the IECEE CB Scheme. Separately from the issuing body requirement, the model name and list of raw materials must be identical to the product applied for, the Annex 2 entry items must be present, and if three years have passed since issue, data confirming no change are additionally required. The cover page of a CB certificate cannot confirm these three things.
- Q. Do test data older than three years have to be retested?
- The provision does not require retesting. It only requires that, for test data issued more than three years before the submission date, data confirming that the product has not changed since the date of issue be submitted in addition. However, if there has been an actual change in the meantime, data confirming no change cannot be produced, so the effect of the change on the existing test results must be reviewed separately.
- Q. In what format should the data confirming no change be submitted?
- The notice does not prescribe the format. The model and model name, the number and date of issue of the reference test data, the period confirmed, and the confirming party and signature should be visible so that the reviewer can connect the test data to the current product. It is safer not to assume that any single form will always be accepted.